GloMax®-20/20 Luminometer
Light Standard
Overview
The GloMax®-20/20 Luminometer Light
Standard provides a quick and easy way to verify the performance of the
GloMax®-20/20 Luminometer. It allows the user to check the reproducibility,
sensitivity, and linearity of the luminometer. It can be used for both
short and long term stability check of the detection system.

The Light Standard consists
of three highly stable light sources that
simulate luminescent samples at three different signal levels.
Operating Instructions
Using the Light Standard on the GloMax®-20/20 Luminometer is very
simple. First, turn OFF the GloMax®-20/20 Luminometer.
Second, insert the Light Standard into the sample compartment. Third,
turn ON the GloMax®-20/20 Luminometer. It will automatically detect
your standard. The Light Standard will report the status of the instrument
in less than 30 seconds.
Dimensions: 2 1/2"x1 5/8'' x 1 3/8''
Weight: 4.7 OZ
Spectral range of light emission: 565±10 nm
Light levels: High 5x107-8x107, Medium 5x106-8x106,
Low 8x105-1.4x106
Operating temperature: 20-35°C
Storage temperature: 10-40°C
Humidity: 5-75%
|