TBS-380 Mini-Fluorometer Method for DNA Quantitation Using Hoechst 33258 Dye 1. INTRODUCTION 2. MATERIALS REQUIRED ·
TBS-380 Mini-Fluorometer (P/N 3800-003) 3. FACTORS TO CONSIDER 3.1 The AT content of a DNA sample affects Hoechst 33258-DNA fluorescence. Hence, it is important to use a standard similar to the samples you are testing. Calf Thymus DNA can often serve as a reference for most plant and animal DNA because it is double-stranded, highly polymerized, and is approximately 58% AT (42% GC). For bacterial DNA, a different standard may be needed because the AT content varies widely depending on the species. 3.2 The conformation (supercoiled, relaxed, circular, linear) of plasmid DNA may result in different Hoechst 33258 binding efficiencies. Thus, it is important to select a standard with similar physical characteristics to your sample. 3.3 Hoechst 33258 fluoresces only about half as much when it binds to single-stranded genomic DNA compared to when it binds to double-stranded genomic DNA. In addition, short pieces of single-stranded DNA will not normally cause Hoechst 33258 to fluoresce in proportion to their concentration. 3.4 Buffers commonly used to extract DNA from whole cells have little or no effect on this assay. 3.5 Low levels of detergent(<0.01% SDS) have little or no effect on this assay. 3.6 Salt concentrations in the sample extract of up to 3 M NaCl do not affect this assay. For peak fluorescence, at least 200 mM NaCl is required for purified DNA, and 2.0 to 3.0 M is required for crude samples. In crude samples, higher salt concentrations appear to cause the dissociation of proteins from DNA, allowing the dye molecules to bind to DNA easier. 3.7 RNA does not interfere significantly with the DNA assay because Hoechst 33258 does not normally bind to RNA. Under high salt concentrations, fluorescence from RNA is usually less than 1% of the signal produced from the same concentration of DNA. 4. SOLUTION PREPARATION NOTE: Hoechst 33258 is a possible carcinogen and possible mutagen. Wear gloves and a mask, and work under a fume hood. 4.1 Hoechst
33258 stock dye solution (1 mg/mL): 4.2 10X TNE
buffer stock solution: *NOTE: The pH and NaCl concentration are essential for proper binding of the Hoechst reagent. 4.3 1X TNE: Dilute 10 mL 10X TNE with 90 mL distilled, 0.45 µm filtered water. 4.4 To prepare
a 2X Dye Solution (200 ng/mL) for 10-1000 ng/mL final DNA concentration: 4.5 Calf
thymus DNA standard: 5. PROTOCOL NOTE: Accurate pipetting and thorough mixing are critical for reproducible results. However, take extreme care when mixing samples; do not introduce air bubbles. Air bubbles can cause scattering of light leading to inaccurate results. If air bubbles form, hold the upper portion of the cuvette in one hand and gently tap the bottom sides of the cuvette with your other hand to release bubbles. 5.1 Choose the assay range most suitable for your samples. For 10 to 1000 ng/mL DNA standards (as shown in Table 1), prepare a series of DNA solution at 2X final concentration. Mix equal volume of the 2X DNA solution with the 2X dye solution prepared in 4.4 into disposable cuvettes or plastic test tubes for transfer to Minicell cuvettes. For blank, mix equal volume of 1X TNE with the 2X Dye Solution. Be sure to add enough volume into the cuvettes. The minimum volume is 2 mL for 10x10mm cuvette and 50 µL for Minicell cuvette
Table 1. DNA standard curve for 10x10 mm cuvette. 5.2 Turn on the fluorometer and select the UV channel. Calibrate the instrument with the highest standard. 5.3 Prepare the unknown samples by diluting samples to desired volume. Add equal volume of the 2X dye solution used for standards and blank. 6. GENERATING A STANDARD CURVE Generating a standard curve verifies the linearity of the assay within a particular concentration range. It is recommended that you perform this at least once when working with a new instrument or performing the assay for the first time. Also, you may want to generate a standard curve every few weeks as a quality check on the standard, a reliability check on the instrument, and a consistency check on technique. NOTE: If
the measured values near one end of the curve deviate consistently from
the line, those values represent a nonlinear region. Sample concentrations
should be adjusted to stay within the linear region of the assay. Orders for
Molecular Probes' products may be placed by: Mailing Address: Information on the scientific and technical back-ground of Molecular Probes' products is available from the Technical Assistance Department by: Phone: (541)
465-8353
Orders for Turner BioSystems' products may be placed by: Phone: (408)
636-2400 Contact us
via our contact
form Mailing Address:
|
|