Modulus Single Tube Luminometer Method for Promega's Steady-Glo® 1. INTRODUCTION The Modulus by Turner BioSystems in combination with Promega's Steady-Glo® Assay kit provides a convenient procedure for quantifying gene expression under the luciferase reporter system. Transcriptional regulation, coupled to the expression of a luciferase reporter gene, is regularly used to study a wide range of biological events in cultured cells. The luminescent signal half-life of the Steady-Glo Reagent is approximately five hours. The light signal can be measured between 5 minutes and several hours after adding the reagent. The Steady-Glo Reagent is compatible with commonly used culture media for mammalian cells (RPMI 1640, MEMa, DMEM and Ham's F12) and tolerates phenol red and organic solvents. In combination
with the Luminescence Module and Modulus, measurements are linear from
1X10-18 to 1X10-11 moles luciferase or orders of
magnitude (Figure 1). All tests were conducted using Steady-Glo Luciferase
Assay Kit (Promega Catalog # E2520) and purified recombinant firefly luciferase
enzyme (Promega Catalog # E1701). 2. MATERIALS REQUIRED
3. EXPERIMENT PROTOCOL 3.1 Reagent
Preparation 3.2 Instrument Setup 3.2.1 Power OFF the Modulus. Install the Luminescence Module according to the Operating Manual. 3.2.1 Power ON the Modulus. A 5-minute countdown immediately after power up allows a warm-up period for the Modulus. 3.2.2 After the warm-up, the Modulus is ready to measure your samples. The default protocol with 1-second integration is suitable for the Steady-Glo Assay. 3.3 Sample Analysis 3.3.1 Remove the cell cultures from the incubator. NOTE: For maximum reproducibility, equilibrate cell cultures to room temperature before adding reagent. 3.3.2 Add a volume of the Steady-Glo Reagent equal to that of the culture medium. 3.3.3 Wait a minimum of 5 minutes to allow for sufficient cell lysis. Then transfer the sample to a 1.5 mL centrifuge tube for analysis. 3.3.4 Insert the tube into the Modulus and touch "Measure Luminescence" to commence measurement. 3.3.5 Record the results in Relative Light Units (RLU). The Modulus displays the most recent 20 results on the touchscreen. 4. ABOUT PROMEGA CORPORATION Steady-Glo is a trademark of Promega Corporation and is registered with the U.S. Patent and Trademark Office. Orders for Promega's products may be placed by: Phone: (800)
356-9526 5. ABOUT TURNER BIOSYSTEMS Modulus is a trademark of Turner BioSystems. Orders for Turner BioSystems' products may be placed by: Phone: (408)
636-2400 Contact us
via our contact
form Mailing Address: CAUTION: The lyophilized Steady-Glo Substrate contains dithiothreitol (DTT) and is therefore classified as hazardous. The reconstituted reagent is not known to present any hazards as the concentration of DTT is less than 1%. However, we recommend the use of gloves, lab coats and eye protection when working with these or any chemical reagents. Promega and Turner BioSystems assume no liability for damage resulting from handling or contact with these products. |
|
| Promega Corporation Sunnyvale Office • Detection Instruments 645 N Mary Avenue • Sunnyvale CA 94085 Phone: 408.636.2400 - Toll Free: 888.636.2401 - Fax: 408.737.7919 |
||
|
||