BCA Protein Assay 1. INTRODUCTION The Modulus by Turner BioSystems and the BCA Protein Assay kit from Pierce provides a simple method for quantifying protein over a large range. The Bicinchoninic acid (BCA) method employs the reduction of Cu+2 to Cu+1 by protein in an alkaline medium. The combination of BCA and Cu+1 creates a purple-colored product that absorbs at 562 nm. The amount of product formed is dependent upon the amount of protein in the sample. Individual proteins differ in their color responses, and one must consider this factor during protein analysis. Temperature, detergents, salts and various buffer components may affect the assay. Refer to the product literature supplied with the BCA Protein Assay kit for more information before performing the assay. The minimum
concentration of Bovine Serum Albumin (BSA) read by the Modulus Absorbance
Module and the BCA Protein Assay is 5 µg/mL (Figure 1) and the maximum
is 2 mg/mL (Figure 2). The Enhanced Protocol requires a 60°C incubation
for 30 minutes. The Standard Protocol permits either a 30-minute incubation
at 37°C or a 2 hour-incubation at room temperature. 2. MATERIALS
3. PREPARATION NOTE: Store kit contents at room temperature. 3.1 BSA Standard Curve 3.4 Instrument Setup 3.4.1 Power OFF the Modulus. Install the Absorbance Module according to the Operating Manual. 3.4.2 Power ON the Modulus and use the touchscreen to confirm that the Absorbance Module is installed. 3.4.3 Touch "Calibrate" and use the black cuvette to set the Modulus to calibrate the zero (dark) reading. 3.4.4 Use a cuvette containing 2 mL of ultrapure water to calibrate the baseline (100% transmittance) reading. 3.4.5 Touch "OK" to accept the calibrations and return to the "Home" screen. 4. SAMPLE ANALYSIS 4.1 Transfer each sample and standard to a 10 x 10 mm cuvette. 4.2 Insert the cuvette into the Modulus and touch "Measure Absorbance" to commence measurement. NOTE: The BCA assay does not reach an end point. At room temperature, the final color of the reaction continues to develop over time at a slow rate. It is advised to read all samples and standards within a 10-minute period. 4.3 Record the results in Absorbance units (Ab). 4.4 Use a standard curve to determine the protein concentration of each unknown sample. A four-parameter (quadratic) or best-fit curve provides the best accuracy. 5. ABOUT PIERCE Orders for
Pierce products may be placed by: Mailing Address: 6. ABOUT TURNER BIOSYSTEMS Modulus is a trademark of Turner BioSystems. Orders for Turner BioSystems' products may be placed by: Phone: (408)
636-2400 Contact us
via our contact
form Mailing Address:
|
|
| Promega Corporation Sunnyvale Office • Detection Instruments 645 N Mary Avenue • Sunnyvale CA 94085 Phone: 408.636.2400 - Toll Free: 888.636.2401 - Fax: 408.737.7919 |
||
|
||