Modulus Single Tube Multimode Reader Method for Coomassie Plus - The Better Bradford Assay Kit 1. INTRODUCTION The Modulus
by Turner BioSystems in combination with the Pierce's Coomassie Plus
Assay kit provides a convenient procedure for quantifying protein. When
the Coomassie dye binds protein the absorption maximum shifts from 465
nm to 595 nm. The Absorbance Module detects as little as 25 µg of
BSA in 1.5 mL of Coomassie Plus Reagent. Figure 1. Bradford Assay was performed on the Modulus using the Absorbance Module, the 600 nm filter, and the Coomassie Plus Reagent. Each point represents the average of replicate samples after background subtraction (n=2). 2. MATERIALS
3. PREPARATION NOTE: Store assay reagent at 4°C. Unopened vials of bovine serum albumin (BSA) standard may be stored at room temperature. 3.1 BSA
Standard Curve 3.2.1 Gently invert the bottle of Coomassie Plus Reagent solution to mix the solution before removing the amount necessary for the assay. 3.2.2 Equilibrate the reagent to room temperature before use. 3.4 Instrument Setup 3.4.1 Power OFF the Modulus. Install the Absorbance Module into the sample compartment according to Operating Manual. 3.4.2 Insert the 600 nm filter paddle in the Absorbance Module. 3.4.4 Touch "Calibrate" and use the black cuvette to set the Modulus to calibrate the zero (dark) reading. 3.4.5 Use a cuvette containing 2 mL of ultrapure water to calibrate the baseline (100% transmittance) reading. 3.4.6 Touch "OK" to accept the calibrations and return to the "Home" screen. 4. SAMPLE
ANALYSIS Mailing Address: 6. ABOUT TURNER BIOSYSTEMS Modulus is a trademark of Turner BioSystems. Orders for Turner BioSystems' products may be placed by: Phone: (408)