A TD-700 Laboratory Fluorometer
Method for
CyQUANT Cell Proliferation Assay
1. Introduction
The Turner BioSystems TD-700 Laboratory
Fluorometer in combination with Molecular Probes' CyQUANT Cell Proliferation
Assay Kit provides a convenient, rapid and sensitive procedure for determining
the density of cells in culture. The assay has a linear detection range
extending from 50 or fewer to at least 200,000 cells in 2 mL volumes (Figure
1) and is therefore ideal for cell proliferation studies as well as for
routine cell counts. The assay is based on dye fluorescence enhancement
upon binding to cellular nucleic acids.1 Cells are simply lysed
by addition of a buffer containing the CyQUANT-GR dye; there are no washing
steps, growth medium changes or long incubations. The resulting fluorescence
is proportional to the number of cells in the sample and is measured directly
using the TD-700 fluorometer equipped with a fluorescein filter kit. The
CyQUANT assay can detect much lower cell numbers than Neutral Red or methylene
blue assays.2,3,4 Unlike procedures that rely on the conversion
of tetrazolium dyes to blue formazan5 products or 3H thymidine
incorporation assays,6 the CyQUANT method is rapid and does
not rely on cellular metabolic activity. Thus, cells can be frozen prior
to assaying; time course assays are facile and data obtained from samples
taken at widely different time intervals can be directly compared. Also,
unlike tetrazolium conversion assays, serum components do not appreciably
interfere with the assay. The CyQUANT assay performs reliably with widely
disparate cell types, including mouse fibroblasts (NIH 3T3 and CREBAG
2 cells), normal human umbilical vein endothelial cells (HUVEC, InvitroCyte,
Inc., Seattle), canine kidney cells (MDCK), chinook salmon embryo cells
(CHSE), rat basophilic leukemia (RBL), rat glioma cells (C6) and mouse
myeloma cells (P3X63A68).
2. Materials Required
- TD-700 Fluorometer with standard PMT
and 10 x 10 mm square cuvette adaptor (P/N 7000-009).
- Blue Mercury Vapor Lamp (P/N 10-089)
or Quartz-halogen lamp (P/N 7000-930).
- Fluorescein filter kit (P/N 10-086R).
- 10 x 10 mm square methacrylate disposable
cuvettes (P/N 7000-959).
- CyQUANT Cell Proliferation Assay
Kit (catalog number C-7026), supplied by Molecular Probes, Inc., Eugene,
OR. The kit contains 550 µL of 400X CyQUANT-GR stock solution in DMSO,
11 mL of 20X Cell Lysis Buffer, and 100 µL of 100 µg/mL lamda -DNA Standard
in TE buffer (10 mM Tris, 1 mM EDTA). The kit contents are sufficient
for 100 assays using 2.0 mL samples in 10 x 10 mm cuvettes. Handling,
storage and the use of the reagents should be performed in accordance
with the product information sheet supplied by Molecular Probes, Inc.
3. Experiment Protocol
3.1 Reagent Preparation
On the day of the experiment, dilute the
concentrated Cell Lysis Buffer stock solution 20-fold in distilled water
(2.0 mL will be required for each assay). Just prior to running the experiment,
dilute the CyQUANT-GR stock solution 400-fold into the 1X Cell Lysis Buffer.
For example, to prepare 20 mL of CyQUANT-GR working solution (enough for
~10 assays), first make the 1X Cell Lysis Buffer by mixing 1 mL of the
20X stock with 19 mL of nuclease-free distilled water; next add 50µL of
the CyQUANT-GR stock solution and mix thoroughly. We recommend preparing
the working solution in a plastic container, rather than in glass; the
CyQUANT-GR reagent may adsorb to glass surfaces. Protect the working solution
from light by keeping it in an opaque bottle, covering it with foil or
placing it in the dark to prevent photodegradation of the CyQUANT-GR dye.
For best results, the solution should be used within a few hours of its
preparation.
3.2 Cell-Number Standard Curve
A reference standard curve can be created
for converting sample fluorescence values into cell numbers. The cell
type used for the standard curve should be the same as that which is used
in the experiment. It is possible to assay either suspension cells or
adherent cells, however, the latter must first be detached and suspended
by treatment with trypsin. Note that some adherent cells are sensitive
to trypsinization and some cell lysis might ensue.
3.2.1 Prepare a concentrated cell suspension
in medium: ideally this should be ~1 mL total volume at a density of about
5 x (10^5) - 1 x (10^6) cells/mL. Determine the actual cell density by
counting the cells using a hemacytometer(7) or optical density measurements.(8)
3.2.2 Centrifuge 1.0 mL of the concentrated cell suspension for 5 minutes
at 200 X g (1500 rpm in a microcentrifuge). Carefully remove and discard
the supernatant without disturbing the cell pellet, and freeze the cell
pellet at -70EC (note [A]).
3.2.3 Thaw the cell pellet at room temperature, add 1.0 mL of the CyQUANT-GR
dye/Cell Lysis Buffer (prepared in Section 3.1), and resuspend the cells
by briefly vortexing.
3.2.4 In a set of culture tubes, serially dilute the cell suspension with
CyQUANT-GR dye/Cell Lysis Buffer to obtain a range of 2.0 mL samples containing
numbers of cells from 50 to 200,000. Include a 2.0 mL sample with no cells
as a reagent blank. Incubate for 2-5 minutes at room temperature, protected
from light.
3.2.5 Transfer samples to standard acrylic fluorescence cuvettes and measure
the fluorescence using the TD-700 fluorometer with the fluorescein filter
set (P/N 10-086R). Insert the most fluorescent sample first and calibrate
the instrument sensitivity as directed in the TD-700 manual (press #2,
calibrate). This procedure automatically optimizes the instrument sensitivity
to match the fluorescence of the sample.
3.2.6 Measure the fluorescence of the remaining samples. To equalize
any photobleaching effects, insert samples into the fluorometer for approximately
equal time periods. The fluorescence value of the reagent blank (CyQUANT-GR
dye + Cell Lysis Buffer only) may be subtracted from that of each sample.
Corrected or uncorrected data may be used to generate a standard curve
of fluorescence versus cell number (for example, see Figure 1). The form
of the standard curve will vary with cell type.

Figure 1. Quantitation of MDCK (Madin-Darby canine kidney) cells
using the CyQUANT Cell Proliferation Assay Kit and the TD-700 fluorometer.
Upper panel: 1000 to 200,000 cells per 2.0 mL sample. Lower panel: 50
to 1000 cells per 2.0 mL sample.
3.3 Cell-Number Determination: Cells
Grown in Standard Culture Conditions
The CyQUANT Cell Proliferation Assay Kit
can be used to count the number of cells in a sample taken from a conventional
cell culture. Adherent cells must be first detached by treatment with
trypsin and then suspended. Note that some lysis may occur in adherent
cells that are sensitive to trypsinization. Cells grown in suspension
can be assayed directly.
3.3.1 Transfer suspended cell samples to
centrifuge tubes and centrifuge for 5 minutes at 200 X g (e.g., 1500 rpm
in a microcentrifuge). Samples should contain from about 50 to about 200,000
cells. Remove and discard the supernatant without disturbing the cell
pellet, and freeze the cell pellet at -70EC (note [A]).
3.3.2 Thaw the cell pellets at room temperature and add 2.0 mL of CyQUANT-GR/Cell
Lysis Buffer (prepared in Section 3.1) to each sample.
3.3.3 Transfer entire 2.0 mL samples to acrylic fluorescence cuvettes
and measure the fluorescence using the TD-700 fluorometer using the same
instrument parameters as used in generating the standard curve (Section
3.2.5). To equalize any photobleaching effects, insert samples into
the fluorometer for similar time periods to those used for the standard
curve measurements.
3.3.4 If the standard curve was plotted using blank-subtracted data
(Section 3.2.5), the reagent blank (CyQUANT-GR dye + Cell Lysis Buffer
only) fluorescence value must also be subtracted from that of each of
the samples. Convert the observed fluorescence to cell number using a
standard curve (Section 3.2.6). We have found that many different cell
types can be assayed using this procedure, but the absolute signal is
cell type-dependent. Thus it is advisable to use a standard curve generated
from the same cell type that is being assayed, for comparison. Alternatively,
a standard curve generated using pure DNA (Section 3.5) can be calibrated
relative to an appropriate cell type.
3.4 Cell Number Determination Based
on DNA or RNA Alone
In the protocols described above, the CyQUANT-GR
reagent is used to determine cell number by staining both DNA and RNA.
DNA to RNA ratios, however, may vary according to cell type and cell cycle
position. Fluorescence due to CyQUANT-GR dye binding to RNA can be eliminated
by pretreating samples with DNase-free RNase. Likewise, fluorescence due
to DNA-bound dye can be eliminated by pretreating samples with RNase-free
DNase.
3.4.1 For determination of total cellular
DNA or RNA, freeze a cell pellet containing 20,000-100,000 cells at -70EC,
thaw at room temperature, and resuspend in 1 mL of 1X Cell Lysis Buffer,
containing 180 mM NaCl. For RNase treatment, this buffer should also contain
1 mM EDTA. For DNase treatment the buffer should contain 1 mM CaCl2 and
1 mM MgCl2.
3.4.2 DNase-free-RNase A or RNase-free-DNase I is added to a final concentration
of about 1.35 Kunitz units/mL (RNase) or 45 Kunitz units/mL (DNase). (9,10)
Samples are incubated for one hour at room temperature.
3.4.3 An equal volume of a 2X working solution of CyQUANT-GR dye (50µL
of the CyQUANT-GR/DMSO stock solution per 10 mL of 1X Cell Lysis Buffer)
is added to each sample. Samples are incubated for 2-5 minutes.
3.4.4 The fluorescence is measured as described above (3.2.5). It is suggested
that controls be run for each digested sample, using the appropriate buffer,
as the presence of salt and divalent cations slightly reduces the slope
of the standard curve.
3.5 DNA Standard Curve
The CyQUANT Cell Proliferation Assay Kit
includes a 100 µg/mL sample of bacteriophage lambda-DNA that can be used
to prepare a standard curve for DNA content. The standard curve can serve
to quantitate cellular DNA, provided that the cell lysates are pretreated
with DNase-free RNase to eliminate the RNA component of the fluorescence
signal (Section 3.4, above). Alternatively, the standard curve can be
used to calibrate the assay for use of the fluorometer at different times
or on different days. Variation in the signal intensity of the standard
curve is directly related to variation that will be observed for assaying
cells on different days, and is instrument-dependent. In a set of culture
tubes, prepare serially diluted 2.0 mL samples of bacteriophage lamda-DNA
with concentrations ranging from 50 pg/mL to 0.5 µg/mL by diluting the
100 µg/mL stock solution into CyQUANT-GR/Lysis Buffer working solution
(as prepared in Section 3.1). Include a reagent blank sample without DNA.
Measure fluorescence as described above (3.2.5 and 3.2.6).
Note
[A] Frozen cell pellets in centrifuge tubes can be stored at -70° C for
up to four weeks. The freezing step is important for efficient cell lysis
in the CyQUANT assay.
4. References
- J Immunol Meth 142, 199 © 1991
- In Vitro Toxicology 3, 219 © 1990
- Biotechnic and Histochemistry 68, 29
© 1993
- Anal Biochem 213, 426 © 1993
- Cancer Res 48, 4827 © 1988
- Exp Cell Res 124, 329 © 1979
- Meth Enzymol 58, 141 © 1979
- BioTechniques 21, 260 © 1996
- J Gen Physiol 24, 15 © 1940
- J Gen Physiol 33, 349 © 1950
5. Patent and Trademark Information
CyQUANT is a trademark of Molecular Probes,
Inc. Materials and methods incorporated in the CyQUANT Cell Proliferation
Assay Kit are covered by current or pending U.S. and foreign patents.
6. About Molecular Probes
Orders for Molecular Probes' products may
be placed by:
Phone: (541) 465-8338 or Toll-Free (800) 438-2209 in the U.S. and Canada
Fax: (541) 344-6504 or Toll-Free (800) 438-0228 in the U.S. and Canada
Email: order@probes.com
Mail: Molecular Probes, Inc., PO Box 22010, Eugene, OR 97402-0469, USA
Information on the scientific and technical
background of Molecular Probes' products is available from the Technical
Assistance Department by:
Phone: (541) 465-8353
Fax: (541) 465-4593
Email: tech@probes.com
Internet: http://www.probes.com
|