A Modulus Microplate Fluorometer Method for Fluorescein Measurement INTRODUCTION
Figure 1. Broad range of fluorescein in serial dilutions, measured by the Modulus Microplate Fluorometer. 200-uL sample per well, n = 5.
Figure
2. Low-concentration range of fluorescein measured by the Modulus
Microplate Fluorometer. 200-uL sample per well, n=5. MATERIALS
REQUIRED
EXPERIMENTAL
PROTOCOL 1. Reagent Preparation Recommendation Make a serial dilution of fluorescein in 100-mM Sodium Borate buffer. Protect the working solution from light by covering it with foil or placing it in the dark. 2. Instrument Set Up
3. Sample Analysis
RESULTS Sensitivity: typically ~2.5 pM Dynamic Range: linearity range was 2 pM - 625,000 pM Minimum Detection Limit*: 0.5 fmol/200 µL (= 2.5 pM), based on 3 STDV of blank controls, n = 24 * Data determined separately and not shown in this application note. CONCLUSION The Modulus Microplate Fluorometer offers superior sensitivity and dynamic range for detection of fluorescein. The Modulus Microplate Fluorometer achieves its superior performance with a dedicated fluorescence detector instead of sharing the detector with other detection modes. The individual optical kit with solid-state optics and a powerful wavelength-matched LED delivers excellent sensitivity and dynamic range. The modular approach of the Modulus Microplate Fluorometer allows for instrument capability expansion as needs in the lab change. Luminescence and/or absorbance detection modules as well as other accessories can be added after the initial purchase. The superior performance, ease of use, and utmost flexibility of the Modulus Microplate make it an ideal microplate reader for today's life science laboratory. ABOUT TURNER BIOSYSTEMS, INC. Modulus is a trademark of Turner BioSystems. Orders for Turner BioSystems' products may be placed by: Phone: (408)
636-2400 Contact us
via our contact form Mailing Address: TRADEMARKS For research use only. Not for use in diagnostic procedures. Modulus is a trademark of Turner BioSystems, Inc. All other trademarks are the sole property of their respective owners. |
| Promega Corporation Sunnyvale Office • Detection Instruments 645 N Mary Avenue • Sunnyvale CA 94085 Phone: 408.636.2400 - Toll Free: 888.636.2401 - Fax: 408.737.7919 |
||
|
||